Model Number 71640634
Medical Device Problem Code Difficult or Delayed Separation (4044)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during surgery the meta-tan 10mm x 34cm right was not allowing a 5.0mm screw to pass through it.The screws got stuck in the nail.The procedure was completed, with a significant delay, using the same nail.Patient was not harmed beyond the problem reported.
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
META-TAN 10MM X 34CM RIGHT
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14520273
Report Number1020279-2022-02641
Device Sequence Number5604145
Product Code HSB
UDI-Device Identifier00885556340196
UDI-Public00885556340196
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K092748
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71640634
Device Catalogue Number71640634
Device Lot Number20LT44280
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2022
Supplement Date Received by Manufacturer06/24/2022
Initial Report FDA Received Date05/27/2022
Supplement Report FDA Received Date06/24/2022
Date Device Manufactured11/18/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1