| Model Number |
UNISG |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported a fracture on the gold tile screw.Doctor had experience the same issue back in 2017 with this patient.Patient is buxist.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmerbiomet complaint number (b)(4).The following sections have been updated: h3: device evaluated by manufacturer: change ¿no' to 'yes.' product evaluation: one certain® gold-tite® square uniscrew, unisg was returned for investigation.Visual inspection via naked eye and camera magnification confirmed that the screw was fractured at the threads.Functional testing to recreate the reported event could not be performed due to the nature of the device and event.Pre-existing conditions were noted on the complaint: bruxism.The reported device was located at tooth site #46 (fdi).Pictures or x-ray images were not provided.Review of appropriate documentation: documents reviewed: biomet 3i restorative products ifu (p-iis086gr) rev f - october 2019.Information identified: 'precautions' 'breakage' 'warning.' per the applicable ifu, breakage may occur when device is loaded beyond its functional capability.Dhr review: dhr review was completed for the subject lot number (1206880).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review: a lot number specific complaint history review was performed for the subject lot number (1206880) for similar events, using complaint keywords : fracture screw.The complaint history review revealed that there are no existing non-conformances/capa/hhe/d/ie/product holds for the reported product.Post market trend review: aug post market trending was reviewed and there were no negative trends or corrective actions for the reported event (fracture screw) and device (unisg).Malf/event: based on the available information, device malfunction did occur and the reported event was confirmed following inspection and evaluation.
|
| |
|
Search Alerts/Recalls
|