We checked the returned unit and confirmed that the ccd module wire was broken.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module wire.In addition, we confirmed that the ccd driver pcb fluid damage, the electrical connector fluid damage, the lg cable connector fluid damage, the lg cable connector corroded, the shield cover corroded, and the insertion flexible tube (ift) crushed; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "(b)(4) (image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
|