• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE (127 DEG) SIZE 5.5 ACCOLADE (127 DEG) SIZE 5.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE (127 DEG) SIZE 5.5 ACCOLADE (127 DEG) SIZE 5.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 6021-5537
Medical Device Problem Codes Corroded (1131); Degraded (1153); Material Erosion (1214)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Metal Related Pathology (4530)
Date of Event 02/07/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his left hip on or about (b)(6) 2015 and was revised on (b)(6) 2022.It is further alleged that he suffered injuries as a result of implantation and explantation of the device(s) at issue, device recall and excessive levels of chromium and cobalt in his blood.
 
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Not available.
 
Additional Manufacturer Narrative
An event regarding abnormal ion level, corrosion and wear (trunnionosis) involving an accolade stem was reported.The events were confirmed via medical review.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-medical records received and evaluation: a review of the provided medical information by a clinical consultant indicated: "the implant sheet confirms that the patient had a cocr v40 femoral head implanted.In the indications portion of the operative report the surgeon notes that pre-operative cobalt and chromium levels were slightly elevated at 3.6 and 0.8 respectively.A pre-operative mri showed a fluid collection and particulate debris.These mri findings were corroborated at the time of surgery.Upon removal of the femoral head some "fouling " was noted on the trunnion.He stated further that this fouling was easily wiped away with a wet sponge and the trunnion was "intact" the patient having been previously implanted with a v40 cocr head is confirmed as is a revision surgery for trunnionosis.The indications section of the operative record reports minimal pre-operative serum metallic ion elevations and a pre-operative mri showed fluid with particulate debris, in the operative report the surgeon describes this fluid and debris as "like wet cat food".With this limited information the root cause of this issue cannot be determined." -product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: a review of the provided medical information by a clinical consultant indicated: "the implant sheet confirms that the patient had a cocr v40 femoral head implanted.In the indications portion of the operative report the surgeon notes that pre-operative cobalt and chromium levels were slightly elevated at 3.6 and 0.8 respectively.A pre-operative mri showed a fluid collection and particulate debris.These mri findings were corroborated at the time of surgery.Upon removal of the femoral head some "fouling " was noted on the trunnion.He stated further that this fouling was easily wiped away with a wet sponge and the trunnion was "intact" the patient having been previously implanted with a v40 cocr head is confirmed as is a revision surgery for trunnionosis.The indications section of the operative record reports minimal pre-operative serum metallic ion elevations and a pre-operative mri showed fluid with particulate debris, in the operative report the surgeon describes this fluid and debris as "like wet cat food".With this limited information the root cause of this issue cannot be determined." no further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his left hip on or about (b)(6), 2015 and was revised on (b)(6), 2022.It is further alleged that he suffered injuries as a result of implantation and explantation of the device(s) at issue, device recall and excessive levels of chromium and cobalt in his blood.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE (127 DEG) SIZE 5.5 ACCOLADE (127 DEG) SIZE 5.5
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
brad curtis
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14520826
Report Number0002249697-2022-00774
Device Sequence Number6530308
Product Code LPH
UDI-Device Identifier04546540510778
UDI-Public04546540510778
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/12/2020
Device Model Number6021-5537
Device Catalogue Number6021-5537
Device Lot Number50456304
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/06/2022
Supplement Date Received by Manufacturer08/04/2022
Initial Report FDA Received Date05/27/2022
Supplement Report FDA Received Date08/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/13/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-