| Model Number |
351-07-04 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during impaction of then talar implant, the talar implant inserter broke.No parts or pieces fell into the patients wound site.There was no surgical delay/prolongation.Patient was last known to be in stable condition following the event.Device is returning.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
The broken talar impactor cover reported was likely the result of additional stresses on the retaining tabs during impaction due to minimal joint space in a mobile bearing, which led to crack initiation, propagation, and ultimate fracture of the device.
|
| |
|