• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. VANTAGE; LP TALAR IMPACTOR INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. VANTAGE; LP TALAR IMPACTOR INSERT Back to Search Results
Model Number 351-07-04
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during impaction of then talar implant, the talar implant inserter broke.No parts or pieces fell into the patients wound site.There was no surgical delay/prolongation.Patient was last known to be in stable condition following the event.Device is returning.
 
Additional Manufacturer Narrative
Pending evaluation.
 
Additional Manufacturer Narrative
The broken talar impactor cover reported was likely the result of additional stresses on the retaining tabs during impaction due to minimal joint space in a mobile bearing, which led to crack initiation, propagation, and ultimate fracture of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VANTAGE
Common Device Name
LP TALAR IMPACTOR INSERT
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key14521104
Report Number1038671-2022-00622
Device Sequence Number6428549
Product Code LXH
UDI-Device Identifier10885862563415
UDI-Public10885862563415
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
UNK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number351-07-04
Device Catalogue Number351-07-04
Device Lot Number145714002
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/10/2022
Supplement Date Received by Manufacturer08/09/2022
Initial Report FDA Received Date05/27/2022
Supplement Report FDA Received Date08/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-