Model Number SESR01
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code Dyspnea (1816)
Date of Event 03/23/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient experienced shortness of breath (sob).The right ventricular (rv) lead exhibited elevated thresholds in bipolar configuration.Initially, the rv lead was reprogrammed to unipolar configuration, and capture management increased unipolar outputs to approximately maximum, triggering recommended replacement time (rrt) in the implantable pulse generator (ipg), triggering mode change, causing the patient's sob.Approximately six weeks later, the rv lead was capped and replaced and the ipg was explanted and replaced. no further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Concomitant medical products: evolutpro-29-us valve, implanted: (b)(6) 2018.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSIA IPG
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key14521802
Report Number3004209178-2022-06855
Device Sequence Number6635452
Product Code NVZ
UDI-Device Identifier00643169708945
UDI-Public00643169708945
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P980035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/14/2019
Device Model NumberSESR01
Device Catalogue NumberSESR01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/29/2022
Initial Report FDA Received Date05/27/2022
Date Device Manufactured02/27/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5076-58 LEAD, 5076-52 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
Patient Weight102 KG
-
-