• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: K2M, INC. SCREW DRIVER W/ THREAD; SIZE 10; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

K2M, INC. SCREW DRIVER W/ THREAD; SIZE 10; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 8801-90034
Medical Device Problem Codes Fracture (1260); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/29/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during removal of a plate two ozark screwdrivers became broken and unusable at the tip.This report captures the second screwdriver.
 
Additional Manufacturer Narrative
Visual inspection was performed.The tip of the driver is fractured.All points of the hexalobe are highly rotated just proximal to the fracture.Complaint history records were reviewed for this catalog number, no adverse trends were identified.It was reported via follow up communication that the or team could not remove a femoral plate.So they suggested using the ozark drivers.The drivers were damaged as they were used off-label on a different system that is not compatible with ozark drivers.The brand and femoral plate type is unknown.The most likely root cause was determined to be off-label use of the drivers to remove femoral plate screws.Ozark drivers are intended for ozark acdf screws only.
 
Event or Problem Description
It was reported that during removal of a plate two ozark screwdrivers became broken and unusable at the tip.This report captures the second screwdriver.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREW DRIVER W/ THREAD; SIZE 10
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key14522142
Report Number3004774118-2022-00210
Device Sequence Number12369358
Product Code LXH
UDI-Device Identifier10888857400863
UDI-Public10888857400863
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8801-90034
Device Catalogue Number8801-90034
Device Lot NumberMPRU
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/02/2022
Supplement Date Received by Manufacturer07/25/2022
Initial Report FDA Received Date05/27/2022
Supplement Report FDA Received Date08/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-