| Model Number |
8801-90034 |
| Medical Device Problem Codes |
Fracture (1260); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during removal of a plate two ozark screwdrivers became broken and unusable at the tip.This report captures the second screwdriver.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection was performed.The tip of the driver is fractured.All points of the hexalobe are highly rotated just proximal to the fracture.Complaint history records were reviewed for this catalog number, no adverse trends were identified.It was reported via follow up communication that the or team could not remove a femoral plate.So they suggested using the ozark drivers.The drivers were damaged as they were used off-label on a different system that is not compatible with ozark drivers.The brand and femoral plate type is unknown.The most likely root cause was determined to be off-label use of the drivers to remove femoral plate screws.Ozark drivers are intended for ozark acdf screws only.
|
| |
|
Event or Problem Description
|
|
It was reported that during removal of a plate two ozark screwdrivers became broken and unusable at the tip.This report captures the second screwdriver.
|
| |
|