MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES 90K¿ IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES 90K¿ IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1400-01
Medical Device Problem Code Expulsion (2933)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Swelling/ Edema (4577)
Type of Reportable Event Serious Injury
Event or Problem Description
The recipient reportedly experienced swelling, pain and device extrusion starting in (b)(6) 2021.The recipient's device was explanted.
 
Additional Manufacturer Narrative
The recipient was given several medications (type unknown) prior to explant.The recipient had an infection that has resolved since the explant surgery.The recipient will be reimplanted at a later date.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Additional Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Additional Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The recipient has reportedly healed.The external visual inspection revealed the electrode array was severed.This is believed to have occurred during revision surgery.The device passed photographic imaging inspection.System lock was verified.The condition of the electrode prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.This device passed all of the tests performed.This version of the hires 90k device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ IMPLANT
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
jennifer rhudy
28515 westinghouse place
valencia, CA 91355
MDR Report Key14522159
Report Number3006556115-2022-00783
Device Sequence Number20465726
Product Code MCM
UDI-Device Identifier07630016816253
UDI-Public(01)07630016816253(11)160909(17)180831
Combination Product (Y/N)N
Initial Reporter CountryEG
PMA/510(K) Number
P960058
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date08/31/2018
Device Model NumberCI-1400-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Not provided
Initial Report FDA Received Date05/27/2022
Supplement Report FDA Received Date06/24/2022
08/29/2022
11/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/09/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-