| Model Number |
7012-90095 |
| Medical Device Problem Codes |
Fracture (1260); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the tips of two mojave pl drivers sheered off intra-operatively.There were no adverse consequences to the patient.The procedure was completed successfully without surgical delay.This report captures the second driver.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been received and fields d4 and d9/h3 have been updated.
|
| |
|
Event or Problem Description
|
|
It was reported that the tips of two mojave pl drivers sheered off intra-operatively.There were no adverse consequences to the patient.The procedure was completed successfully without surgical delay.This report captures the second driver.
|
| |
|
Additional Manufacturer Narrative
|
|
It was initially reported that the tips of two mojave drivers sheered off intra-operatively.However, upon receipt of the devices they were actually everest rod reducers.Section d has been updated to reflect the correct device information.Inspection of the device and additional review of the events indicates this incident to be non-reportable.The event caused no serious deterioration in the state of health of a patient or user.Review of product and complaint history indicates that no prior similar incidents have led to a serious deterioration in the past, nor is a recurrence of this incident likely to lead to a serious deterioration in the future.
|
| |
|
Event or Problem Description
|
|
It was initially reported that the tips of two mojave drivers sheered off intra-operatively.However, upon receipt of the devices they were actually everest rod reducers.There were no adverse consequences to the patient.The procedure was completed successfully without surgical delay.This report captures the second driver.
|
| |
|