SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH BN PROSPEC SYSTEM Back to Search Results
Model Number BN PROSPEC SYSTEM
Medical Device Problem Code Smoking (1585)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An outside of the united states customer contacted a siemens customer care center (ccc).A siemens customer service engineer (cse) was dispatched to the site and confirmed that the bn prospec power supply had failed.The cse replaced the power supply, after which the system returned to normal operation.The cause of the event was the power supply failure.The system is performing according to specifications.No further evaluation of this device is required.
 
Event or Problem Description
After the customer turned on their bn prospec system, the system made a banging noise, after which smoke was emitted from the system and vapor filled the room.The system was powered down and was taken out of operation.The system was down for one day and there was a delay in patient testing of one day.There are no known reports of adverse health consequences due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BN PROSPEC SYSTEM
Common Device Name
BN PROSPEC SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
emil-von-behring- str. 76
marburg, D-350 41
GM  D-35041
Manufacturer (Section G)
BIT ANALYTICAL INSTRUMENTS GMBH
registration number: 300360107
am kronberger hang 3
schwalbach, 65824
GM   65824
Manufacturer Contact
christopher aebig
511 benedict ave.
tarrytown, NY 10591
9144153450
MDR Report Key14523926
Report Number9610806-2022-00039
Device Sequence Number5604455
Product Code JZW
UDI-Device Identifier00630414002736
UDI-Public00630414002736
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K001647
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBN PROSPEC SYSTEM
Device Catalogue Number10465217
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/06/2022
Initial Report FDA Received Date05/27/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1