The subject device is expected to be returned to olympus for evaluation.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.The subject device is expected to be returned to olympus for evaluation.
|
|
The customer reported to olympus, approximately two (2) hours into a therapeutic hysterectomy procedure, the ultrasonic bipolar generator (usg-410) displayed two (2) probe damage error codes, u504 and u601, and started beeping.The thunderbeat device was removed from the patient and the probe jaws were cleaned with wet gauze, the electrode was tested, the cable was checked and it was confirmed there was no contact with metal.After performing troubleshooting, the customer reported the system passed testing.The physician proceeded to insert the thunderbeat probe back into the patient¿s abdomen.At the first activation of the thunderbeat, the jaws broke inside the patient.The surgeon immediately removed the thunderbeat device and the broken piece from the abdomen.There was no patient or user harm reported due to the event.The intended procedure was successfully completed with the same usg-410, but with a different thunderbeat.It is unknown how the broken piece was retrieved and if there was a procedural delay.This event includes two reports: this report is for the ultrasonic bipolar generator, usg-410, s/n (b)(4), patient identifier (b)(6).The thunderbeat, tb-0535fcs, lot kr145443, is being submitted under the medwatch with patient identifier (b)(6).
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and correction to g2.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was not returned for evaluation.Based on the results of the investigation, it¿s likely the reported event was attributed to the broken thunderbeat.The root cause of this event was unable to be identified.The following is included in the instructions for use: ¿troubleshooting: if any irregularity is observed with the equipment, follow the below steps.1.Immediately stop the procedure and withdraw any instrument from inside the body cavity.2.Check that the probe tip and the grasping section are intact.3.Check if any error window is displayed.4.If an error window is displayed, first read section, ¿error screen¿ and section, ¿troubleshooting guide¿ and take proper remedial actions according to ¿what to do when an error code is displayed¿.5.If no error window is displayed, take proper remedial actions according to ¿what to do when no error code is displayed¿.¿ olympus will continue to monitor field performance for this device.
|