DEVICOR MEDICAL PRODUCTS INC HYDROMARK BREAST BIOPSY SITE MARKER; BIOPSY SITE IDENTIFIER Back to Search Results
Catalog Number 4010-03-09-T1
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 05/03/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The hydromark biopsy site identifier is a sterile, single use device intended for use after an open surgical or percutaneous breast biopsy procedure to mark the biopsy site.The instruction for use and deployment guide, provided with the device, must be followed to avoid the applicator from shearing.In accordance with iso 13485:2016 and iso 10993-1 requirements, the mammomark¿ biopsy site identifier devices have been tested for biocompatibility for their intended use and patient contact duration per iso 10993-1.The collagen marker and titanium clip are the only mammomark components classified as having permanent patient contact." components of the applicator are classified as "limited patient contact" and qualified for the intended use during the procedure.Applicator materials are considered medical-grade but have not been evaluated for permanent implantation.As with any implanted device, the implant site should be monitored for any signs of irritation or reaction following the surgical procedure.It was reported that a tip shear had occurred with part of the marker left in the breast.The patient was made aware that the marker piece will be surgically removed and will undergo surgery.To date we have not received any information that the spring left behind in the patient has caused a life-threatening disease, permanent impairment of a body function or permanent damage to a body structure, or a condition that necessitates an unexpected medical or surgical intervention to prevent such a disease, disorder or abnormal physical state or permanent impairment or damage.To date we have not been updated as to whether or not the patient has undergone surgery to get the piece of marker removed or not.Mammotome has made 3 attempts for further information regarding the alleged incident.No further information has been provided by the customer.If new information becomes available the incident will be reassessed accordingly.
 
Event or Problem Description
It was reported by the sales representative that dr.(b)(6) sheared marker off into breast of patient.He used this marker down back of revolve 10g needle.The clinic and doctor were made aware that it would need to be removed from patient immediately.Patient was unaware of what happened.Dr.(b)(6) has been advised to let the patient know that the spring would need to be surgically removed.I have let the facility know that the sheared off spring needs to be surgically removed from the patient.At the time i was made aware of the event, the patient was still unaware of what had actually happened.She was not harmed during the actual biopsy procedure.They were going to inform her what had happened and that a surgery would need to occur to remove the sheared off tip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYDROMARK BREAST BIOPSY SITE MARKER
Common Device Name
BIOPSY SITE IDENTIFIER
Manufacturer (Section D)
DEVICOR MEDICAL PRODUCTS INC
300 e business way fifth floor
cincinnati OH 45241
Manufacturer (Section G)
DEVICOR MEDICAL PRODUCTS DE MEXICO
sor juanaines de la cruz
#20152 4-b
parque industrial, tijuana 22440
MX   22440
Manufacturer Contact
vivian rodriguez
300 e business way fifth floor
cincinnati, OH 45241
MDR Report Key14525100
Report Number3008492462-2022-00007
Device Sequence Number12212094
Product Code NEU
UDI-Device Identifier10841911100130
UDI-Public(01)10841911100130(17)240319(10)F12123609D
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K121113
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/19/2024
Device Catalogue Number4010-03-09-T1
Device Lot NumberF12123609D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/03/2022
Initial Report FDA Received Date05/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-