• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Medical Device Problem Codes Defibrillation/Stimulation Problem (1573); Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/09/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer contacted stryker to report a non-critical issue with their device.Upon evaluation of the customer¿s device, stryker observed that the device intermittently displayed "connect electrodes".When the device was able to charge successfully for defibrillation, intermittently it was disarm itself and the shock was not delivered.In this state, defibrillation therapy may be delayed or prevented from being delivered to the patient if needed.In addition it was observed that one shock was below the normal tolerance.As a result, the wrong defibrillation therapy would be delivered, if it were necessary.There was no patient use associated with the reported event.
 
Additional Manufacturer Narrative
Stryker performed an initial evaluation of the customer¿s device and verified the reported issue.Stryker continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Additional Manufacturer Narrative
Stryker further evaluated the customer's device and was unable to duplicate the reported issue.During visual inspection it was observed that the device therapy connector had a pinch mark on one of the therapy connector's high voltage wires.The therapy connector assembly was replaced as a precaution.After other unrelated repairs were completed, proper device operation was observed through functional and performance testing,.The device was returned to the customer for use.Stryker further evaluated the removed therapy connector assembly and confirmed the pinched wire, however the connector was functional.The cause of the reported issue could not be determined.
 
Event or Problem Description
The customer contacted stryker to report a non-critical issue with their device.Upon evaluation of the customer¿s device, stryker observed that the device intermittently displayed "connect electrodes".When the device was able to charge successfully for defibrillation, intermittently it was disarm itself and the shock was not delivered.In this state, defibrillation therapy may be delayed or prevented from being delivered to the patient if needed.In addition it was observed that one shock was below the normal tolerance.As a result, the wrong defibrillation therapy would be delivered, if it were necessary.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key14525174
Report Number0003015876-2022-00807
Device Sequence Number6008837
Product Code MKJ
UDI-Device Identifier00883873911457
UDI-Public00883873911457
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001268
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/09/2022
Supplement Date Received by Manufacturer08/29/2022
Initial Report FDA Received Date05/27/2022
Supplement Report FDA Received Date08/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/27/2011
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-