• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SURGICAL SPECIALTIES MEXICO S. DE. R.L. DE C.V. STRATAFIX; 2CT-1 #2 PDO 24 X 24

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SURGICAL SPECIALTIES MEXICO S. DE. R.L. DE C.V. STRATAFIX; 2CT-1 #2 PDO 24 X 24 Back to Search Results
Model Number SXPD2B402
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A batch review of the finished good lot and components confirmed there were no quality issues noted throughout the incoming inspection, manufacturing, sterilization, in-process, or final inspection process.The needle is an ethicon supplied component.The exact device was not returned for review.Samples from the same lot were not available for review.The third party evaluation report has not been received to date.If samples or the report become available at a later date, the samples or information will be reviewed and added to the file and a follow-up report will be filed.The bending, fracturing, breaking of a needle can occur when needles are gripped with a needle holder, forceps, surgical instrument on or near the swaged area or near the tip of the device, when excessive force is applied, when the device(s) are used in applications involving tortuous tissue or with a needle tip design that may not be appropriate for the specific tissue or procedure.The ¿precautions¿ section in the ifu for the device states, ¿care should be taken to avoid damage when handling.Avoid crushing or crimping the suture material with surgical instruments such as needle holders and forceps.To avoid damaging needle points and swage areas, grasp the needle in an area one-third to one-half of the distance from the swaged end to the point¿.Without receiving ethicon¿s third party analysis, testing sterile devices from the same finished good lot, reviewing magnified photos of the broken device or receiving details regarding the tools utilized to grasp the needle component/size of trocar utilized, procedure performed or the surgeon¿s technique a definitive root cause cannot be determined at this time.
 
Event or Problem Description
Breakage needle.It was initially reported (b)(6) 2022 that the needle broke when sewing during myomectomy procedure at the second stitch.Changed to another one to complete the surgery.There is no report of patient injury.Additional information was later received (b)(6) 2022 indicating intervention.It was reported that the needle had fallen into the body and was retrieved with a needle holder; no injury or further incident reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATAFIX
Common Device Name
2CT-1 #2 PDO 24 X 24
Manufacturer (Section D)
SURGICAL SPECIALTIES MEXICO S. DE. R.L. DE C.V.
corredor tijuana rosarito 2000
24702 b, ejido francisco villa
tijuana 22235
MX  22235
Manufacturer (Section G)
SURGICAL SPECIALTIES CORPORATION
corredor tijuana-rosarito 2000 #24702b
ejido francisco villa
tijuana 22235
MX   22235
Manufacturer Contact
luis giannangelo
247 station drive
suite ne1
westwood, MA 02090
MDR Report Key14525256
Report Number3010692967-2022-00017
Device Sequence Number11969996
Product Code NEW
UDI-Device Identifier10705031207882
UDI-Public(01)10705031207882(20)12(17)260131(10)AAHH278
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source User Facility,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/27/2022
Report Date (Section F) 05/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2026
Device Model NumberSXPD2B402
Device Lot NumberAAHH278
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date04/07/2022
Device Age14 MO
Date Report to Manufacturer05/17/2022
Type of Report(Section G)Fifteen-Day
Initial Date Received by Manufacturer 05/17/2022
Initial Report FDA Received Date05/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-