| Model Number |
68LP-275 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/07/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported, right side deflation.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: analysis of the returned device identified, opening curved on posterior, creases fold, wear abrasion and white particles inner device.Leak test and microscopic analysis was performed which identified, opening crease smooth on posterior.The fill test inspection was performed, the result is no blockage.Based on the device analysis, the final assessment is: an opening crease smooth on posterior side assessed, as fold flaw opening.
|
| |
|