• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. (AF-ST. PAUL) 3 LESION NT1100¿ PAIN MANAGEMENT RF GENERATOR; GENERATOR, LESION, RADIOFREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. (AF-ST. PAUL) 3 LESION NT1100¿ PAIN MANAGEMENT RF GENERATOR; GENERATOR, LESION, RADIOFREQUENCY Back to Search Results
Model Number RFG-NT-1100
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/05/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Further information requested, but not yet received.
 
Event or Problem Description
It was reported the procedure was abandoned because the generator lost power during the procedure and could not be resolved.
 
Additional Manufacturer Narrative
The reported issue was not confirmed.Based on the information provided to abbott and the investigation performed, the root cause of the field reported event cannot be conclusively determined as no abnormalities were identified.The returned generator functioned as anticipated during the evaluation period.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.No capa is required; the root cause of the reported issue could not be confirmed to be a manufacturing, design, or quality system related occurrence.The reported issue will continue to be monitored for trends and reviewed in qdr (quality data review).During processing of this incident, attempts were made to obtain complete patient information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3 LESION NT1100¿ PAIN MANAGEMENT RF GENERATOR
Common Device Name
GENERATOR, LESION, RADIOFREQUENCY
Manufacturer (Section D)
ST. JUDE MEDICAL, INC. (AF-ST. PAUL)
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC. (AF-ST. PAUL)
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key14526734
Report Number2184149-2022-00137
Device Sequence Number5746995
Product Code GXD
UDI-Device Identifier05415067022370
UDI-Public05415067022370
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K052878
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRFG-NT-1100
Device Catalogue NumberRFG-NT-1100
Device Lot Number7140662
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2022
Supplement Date Received by Manufacturer06/29/2022
Initial Report FDA Received Date05/27/2022
Supplement Report FDA Received Date07/17/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/13/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-