Model Number 22042 |
Device Problem
No Audible Alarm (1019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/03/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been returned to the investigation facility to allow for an analysis to be performed.If the product is received for evaluation or new information is obtained, a follow up report will be submitted.
|
|
Event Description
|
The customer reported the device speaker is defective, no tone audible.No patient impact or consequences were reported.
|
|
Event Description
|
The customer reported the device speaker is defective, no tone audible.No patient impact or consequences were reported.
|
|
Manufacturer Narrative
|
Additional manufacturing narrative: other text: the returned device was evaluated.The device powered on and off using both battery and ac power.The speakers were not audible.During internal inspection, the customer's system board was probed and a component of the speaker driver circuit was measured open.The speakers were not audible due to the defective component of the system board.A service history record review reveals that this unit was in the field for over six (6) years with no previous reported issues related to this reported event.
|
|