ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; DEFIBRILLATION LEAD Back to Search Results
Model Number 7122Q/58
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Low impedance (2285); Naturally Worn (2988)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/16/2022
Event Type  Injury  
Event Description
It was reported that oversensing, noise as well as low pacing impedance was observed on the right ventricular (rv) lead.The rv lead was being monitored.There were no patient consequences.
 
Event Description
Additional information was received indicating the right ventricular (rv) lead was explanted and replaced to resolve the event.Subclavian crush was observed on the rv lead.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
Type of Device
DEFIBRILLATION LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key14532061
MDR Text Key292985921
Report Number2017865-2022-11248
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734503174
UDI-Public05414734503174
Combination Product (y/n)Y
Reporter Country CodeIT
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/28/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2020
Device Model Number7122Q/58
Device Catalogue Number7122Q-58
Device Lot NumberA000040929
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/31/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ELLIPSE DR CD2377-36QC.
Patient Outcome(s) Required Intervention;
-
-