BOSTON SCIENTIFIC CORPORATION ORIGEN MINI ICD VR; IMPLANTABLE DEVICE Back to Search Results
Model Number D001
Medical Device Problem Codes High impedance (1291); Pacing Problem (1439); Incorrect, Inadequate or Imprecise Result or Readings (1535); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code Electric Shock (2554)
Date of Event 05/03/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient implanted with this implantable cardioverter defibrillator (icd) system presented to the hospital due to receiving anti-tachycardia pacing (atp) and shocks in the ventricular tachycardia (vt) zone.At the time of the last delivered shock the shock impedance was noted to have reached out of range reaching greater than 125 ohms.All other shocks had recorded impedance measurements within normal limits.Device data was sent to technical services (ts) for review.Device data determined the device recorded a code 1005 associated with an open circuit at the time shock was delivered.This system remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORIGEN MINI ICD VR
Common Device Name
IMPLANTABLE DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14536966
Report Number2124215-2022-17286
Device Sequence Number5538281
Product Code LWS
UDI-Device Identifier00802526533723
UDI-Public00802526533723
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P960040/S306
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/23/2020
Device Model NumberD001
Device Catalogue NumberD001
Device Lot Number109363
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/03/2022
Initial Report FDA Received Date05/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/28/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-