• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ PANOPTIX TORIC UV ABSORBING TRIFOCAL IOL; LENS, MULTIFOCAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ PANOPTIX TORIC UV ABSORBING TRIFOCAL IOL; LENS, MULTIFOCAL Back to Search Results
Model Number TFAT40
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code Visual Impairment (2138)
Date of Event 04/18/2022
Type of Reportable Event Serious Injury
Event or Problem Description
A healthcare professional reported an explant has been planned due to no longer can see clear out of os, feels always blurry and not clear.The lens was off positioned and shifted upwards, therefore to attain the correct vision the lens was altered with more astigmatism then wanted at a +1.50.Additional information has been requested.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information provided in a.2., a.3., b.3., b.5., b.6., b.7., d.10 and h.3.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information has been requested and received which states a planned explant has been scheduled on (b)(6) 2022 with the same lens diopter.
 
Additional Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information has been requested and received which states that the patient only needed a repositioning of the iol, not an explant.
 
Additional Manufacturer Narrative
Additional information provided in h.3., h.6.And h.10.The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.Information was provided that a qualified cartridge was used.The associated handpiece and viscoelastic information was not provided.It is unknown if a qualified handpiece and viscoelastic were used.The product investigation could not identify a root cause for the reported complaint.The lens was repositioned.No further information has been provided.If additional information were to become available, the file will be reopened and updated.There are no other complaints in this lot.Based on our current tracking, there are no adverse trends for this reported complaint.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ PANOPTIX TORIC UV ABSORBING TRIFOCAL IOL
Common Device Name
LENS, MULTIFOCAL
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14538688
Report Number1119421-2022-01153
Device Sequence Number20209473
Product Code MFK
UDI-Device Identifier00380652419208
UDI-Public00380652419208
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P040020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 08/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Model NumberTFAT40
Device Catalogue NumberTFAT40.185
Device Lot Number12720802
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/06/2022
Supplement Date Received by Manufacturer05/26/2022
06/28/2022
07/31/2022
Initial Report FDA Received Date05/29/2022
Supplement Report FDA Received Date06/23/2022
07/22/2022
08/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/27/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D.
Outcome Attributed to Adverse Event Other;
Patient Age65 YR
Patient SexMale
-
-