Model Number MI1200 SYNCHRONY PIN
Medical Device Problem Codes Impedance Problem (2950); Output Problem (3005)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 05/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
In situ testing revealed affected channels.The user reported not using the device since sound quality was poor and a few months ago he experienced some pain around the magnet site.It was suggested an appointment be made with ent and an x-ray to check the implant site.
 
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Additional Manufacturer Narrative
Additional information: according to the currently available information, damage to the active electrode as might be caused by an external mechanical impact appears likely.However, to determine an exact root cause a device investigation of the explanted device is necessary.
 
Event or Problem Description
In situ testing revealed affected channels.The user had stopped wearing the device due to poor sound quality and pain around the magnet site.An x-ray has been recommended, however no date has been set yet.
 
Event or Problem Description
The user_s hearing performance with the device was affected.He stopped wearing the audio processor due to poor sound quality and pain around the magnet site.The user was re-implanted with a new device on (b)(6) 2023.
 
Additional Manufacturer Narrative
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.In addition a reduced electrical impedance between coil and eap/rg (evoked action potentials/ remote ground) electrode was discovered.Furthermore, it was reported that the recipient suffered from facial nerve stimulation since activation.A contribution of the lower-ohmic connection between coil and reference electrode cannot be excluded nor confirmed.This is a final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHRONY
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key14539363
Report Number9710014-2022-00363
Device Sequence Number17395317
Product Code MCM
UDI-Device Identifier09008737094010
UDI-Public(01)09008737094010
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/29/2019
Device Model NumberMI1200 SYNCHRONY PIN
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/25/2022
Supplement Date Received by Manufacturer05/25/2022
05/25/2022
Initial Report FDA Received Date05/30/2022
Supplement Report FDA Received Date06/27/2022
04/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/29/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age21 YR
Patient SexMale
-
-