| Model Number |
MI1200 SYNCHRONY PIN |
| Medical Device Problem Codes |
Impedance Problem (2950); Output Problem (3005)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
05/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
In situ testing revealed affected channels.The user reported not using the device since sound quality was poor and a few months ago he experienced some pain around the magnet site.It was suggested an appointment be made with ent and an x-ray to check the implant site.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: according to the currently available information, damage to the active electrode as might be caused by an external mechanical impact appears likely.However, to determine an exact root cause a device investigation of the explanted device is necessary.
|
| |
|
Event or Problem Description
|
|
In situ testing revealed affected channels.The user had stopped wearing the device due to poor sound quality and pain around the magnet site.An x-ray has been recommended, however no date has been set yet.
|
| |
|
Event or Problem Description
|
|
The user_s hearing performance with the device was affected.He stopped wearing the audio processor due to poor sound quality and pain around the magnet site.The user was re-implanted with a new device on (b)(6) 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.In addition a reduced electrical impedance between coil and eap/rg (evoked action potentials/ remote ground) electrode was discovered.Furthermore, it was reported that the recipient suffered from facial nerve stimulation since activation.A contribution of the lower-ohmic connection between coil and reference electrode cannot be excluded nor confirmed.This is a final report.
|
| |
|