| Model Number |
N/A |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Codes |
Hypersensitivity/Allergic reaction (1907); Pain (1994)
|
| Date of Event |
05/02/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant products: unknown oxford femoral component; catalog#unknown; lot#unknown; unknown oxford bearing; catalog#unknown; lot#unknown.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 3002806535-2022-00245, 3002806535-2022-00247.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a revision surgery approximately 8 months after the implantation due to pain and metal allergy.Additional information requested.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow up report is being submitted to relay additional information.D10: medical product: oxf twin-peg cmntd fem sm pma: catalog#161468, lot#681550; oxf anat brg lt sm size 3 pma: catalog#159540, lot#335150.Visual and dimensional evaluations of the returned product found the device to exhibit some scratches which was likely caused by the explantation process and handling following this.Device history record was reviewed and no discrepancies related to the reported event were found.The likely condition of the device when it left zimmer biomet is conforming to specification.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information at time of this report.Due diligence has been completed and no additional information has been made available.
|
| |
|