| Model Number |
EICA8898-01 |
| Medical Device Problem Codes |
Crack (1135); Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during set-up, the tip of the precise collator wand was cracked and leaking saline.The procedure was completed without a delay using a back-up device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H10: h3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A visual inspection of the returned instrument shows no manufacturing abnormalities.The wand has a crack between the shaft and the handle.The wand cable and connector are free from any physical damage.Product was out of the original packaging.No packaging returned.The device was plugged into the controller and registered settings (7,3).The wand was able to generate plasma and coagulation.There was an observed leak where the wand is cracked.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The complaint was confirmed and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include: (1) excessive force.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|