Model Number 420230-10
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 04/08/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a da vinci-assisted low anterior resection procedure, there was a non-intuitive motion issue with the large clip applier instrument when closing of the grips was attempted.The movement was described as ¿uncontrolled.¿ it was noted that the patient sustained bleeding from the inferior mesenteric artery (ima), but it was not severe.The volume of blood loss was reported to be less than 750 ml and the patient was not transfused with blood.The procedure was completed with the da vinci surgical system.On (b)(6) 2021, intuitive surgical, inc.(isi) received the following additional information from the surgeon via a nurse at the site: there was no bleeding in the patient for the procedure and no loose cable seen on the instrument.According to the surgeon, there was uncontrolled movement during application of the clip when closing the jaws.There was no loose cable seen in the instrument during application of the clip.The surgeon checked that the clip was properly latched on after application of the clip.The procedure was completed robotically, and the procedure was not prolonged due to the issue.There was no fragment that fell into the patient as well as no bleeding that occurred in the patient.
 
Additional Manufacturer Narrative
Based on the information provided, the cause of the reported complication cannot be determined as there are contradictory statements regarding the occurrence of bleeding.The instrument was received, and isi performed a device evaluation.The instrument passed the recognition and engagement tests.The instrument moved intuitively with full range of motion in all directions.The grips opened and closed properly.A clip test was performed, and the instrument passed.The instrument was fully functional.A review of the logs found no failures recorded that were related to the reported complaint.There was no problem detected.No cables were found loose or broken on the distal or proximal end of the instrument.A video clip was provided from this reported procedure and evaluated by an isi clinical development engineer (cde).The investigation indicated that while the video quality was poor, there was unexpected yaw motion during the second clip application.No bleeding was observed.This complaint is being reported due to the following conclusion: during a da vinci-assisted low anterior resection procedure, it was alleged that there was a non-intuitive motion issue with the large clip applier instrument and that bleeding occurred from the ima.While the bleeding was noted to not be severe, did not require a blood transfusion, and is not currently connected to the report of non-intuitive motion, the reported failure mode could likely cause or contribute to an injury if it were to recur.The cause of the operative complication is unknown.It is also unclear what medical intervention, if any, was administered due to the intra-operative complication.Blank mdr fields: follow-up was attempted, but the missing patient information was either unknown, unavailable, not provided, or not applicable.The expiration date is not applicable.Field is blank because the product is not implantable.Information for the blank fields is not available.Fields are not applicable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST;DAVINCI SI
Common Device Name
LARGE HEM-O-LOK CLIP APPLIER
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key14541361
Report Number2955842-2022-11911
Device Sequence Number12446718
Product Code NAY
UDI-Device Identifier00886874111680
UDI-Public(01)00886874111680(10)N10190408
Combination Product (Y/N)N
Initial Reporter CountryDA
PMA/510(K) Number
K050369
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 04/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number420230-10
Device Catalogue Number420230
Device Lot NumberN10190408 844
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/23/2021
Initial Report FDA Received Date05/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/05/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient Age73 YR
Patient SexFemale
Patient Weight76 KG
-
-