|
It was reported that on the literature review "thermal capsulorrhaphy for isolated posterior instability of the glenohumeral joint without labral detachment"; after a thermal capsulorrhaphy procedure where a capsure coblation wand was used, 1 patient had a bilateral posterior instability.It was treated with bilateral thermal capsulorrhaphy with a month between shoulders and was not immobilized after surgery.2 months after this, the patient experienced recurrent instability, then he noted a marked diminution in subluxation episodes and returned to sports activity.No further information is available.
|
|
H10: h3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.Insufficient product identification information was provided and thus a manufacturing record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states that the images provided in the article have been interpreted within the text; therefore, no further analysis of the images is required.The root cause and/or patient outcome beyond that which was documented in the article could not be confirmed nor concluded; therefore, no further medical assessment is warranted at this time.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|