MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number ATG120144
Medical Device Problem Codes Leak/Splash (1354); Material Rupture (1546); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one atlas gold pta dilatation catheter was returned for evaluation.No specific anomalies noted during the visual evaluation.On the functional test of the device, it was connected to an in-house presto device and the balloon was inflated and it was noted the water was leaking from the distal tip.Further, the balloon was cut and under microscopic observation, it was noted a partial circumferential break was noted to the guide-wire lumen just proximal to the glue bullet.No other functional testing was performed.Therefore, the investigation for the reported balloon rupture remains inconclusive as no evidence could not be noted due to break at the inner guide wire lumen.The investigation was also confirmed for identified leak as water was noted to be leak from the distal tip during the functional testing.The investigation was confirmed for identified catheter tear as a partial circumferential break was noted to the guide-wire lumen just proximal to the glue bullet.A definitive root cause for the reported balloon rupture and identified catheter tear and leak could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.(expiry date: 09/2024).
 
Event or Problem Description
It was reported that during an angioplasty procedure in the iliac vein, the pta balloon allegedly failed to open in the iliac vein upon attempting to inflate.It was further reported that physician removed the balloon from patient's body and identified leakage of the balloon at the proximal end.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS GOLD
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
FUTUREMATRIX INTERVENTIONAL
1605 enterprise street
athens 75751
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14546145
Report Number2020394-2022-00425
Device Sequence Number6736927
Product Code DQY
UDI-Device Identifier00801741060762
UDI-Public(01)00801741060762
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K181323
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberATG120144
Device Catalogue NumberATG120144
Device Lot Number93WF0121
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/20/2022
Initial Report FDA Received Date05/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age31 YR
Patient SexMale
-
-