SYNTHES GMBH 2.7/3.5MM DEPTH GAUGE 0 TO 60MM; GAUGE, DEPTH Back to Search Results
Model Number 03.133.080
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/03/2022
Event Type  malfunction  
Event Description
The tip of the depth gauge broke while being used intra-operatively.A fragment was generated that required additional x-ray imaging and operative time to remove.This report is for one (1) 2.7/3.5mm depth gauge 0 to 60mm.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: part: 03.133.080, synthes lot: h889463, supplier lot: (b)(4), release to warehouse date: 19 june 2020, 10 july 2020, supplier: (b)(4).No nonconformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Photos of the item were received also.Visual analysis of the returned sample revealed that 2.7/3.5mm depth gauge 0 to 60mm had the needle broken and fragment was not received.However; the fragment is visible in the photo.A dimensional inspection for the 2.7/3.5mm depth gauge 0 to 60mm was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the 2.7/3.5mm depth gauge 0 to 60mm would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed the following source controlled drawings reflecting the current and manufactured revisions were reviewed: -2.7 / 3.5 depth gauge for screws 0-60 mm.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7/3.5MM DEPTH GAUGE 0 TO 60MM
Type of Device
GAUGE, DEPTH
Manufacturer (Section D)
SYNTHES GMBH
1302 wright lane east
west chester PA 19380
Manufacturer (Section G)
SYNTHES MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
MDR Report Key14547351
MDR Text Key300786009
Report Number8030965-2022-03572
Device Sequence Number1
Product Code HTJ
UDI-Device Identifier10886982269690
UDI-Public(01)10886982269690
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number03.133.080
Device Catalogue Number03.133.080
Device Lot NumberH889463
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/19/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age67 YR
Patient SexFemale
-
-