| Model Number |
TMMWB13 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Osteolysis (2377)
|
| Date of Event |
04/13/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimvie complaint (b)(4).Patient weight is not provided / unknown.Initial reporter¿s title is not provided / unknown.Additional 510(k) numbers are k113753 and k112160.A summary investigation has been completed for bone loss events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss have not identified any signals indicating potential non-conformance's affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the implant at tooth sites #45 was removed due to medical bone loss.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to report additional information.The following sections are being reported: d4: expiration date and unique identifier (udi) number.D9: device availability.H2: if follow-up, what type.H3: device evaluated by manufacturer.H4: device manufacturer date.H6: type of investigation.H10: additional narratives/data.Dhr review was completed for the subject lot number.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.As documented in the summary investigation, contributing factors for the reported event likely exist outside of zimmer biomet control, including those related to medical conditions / patient habits and surgical technique.Based on the summary investigation, no malfunction occurred upon investigation and no association between the dental implant and bone loss was found that can explain these events.Therefore, the reported event remains non-verifiable as it is a medical condition.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimvie will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No additional or corrected information to report.
|
| |
|
Search Alerts/Recalls
|