| Model Number |
N/A |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Bacterial Infection (1735)
|
| Date of Event |
02/04/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that patient underwent a right hip revision approximately 1 month post implantation due to infection.The head and liner were removed and replaced.Attempts have been made and additional information on the reported event is unavailable at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant medical products: 110010245 ¿ g7 shell ¿ 65048747; 00625006525 ¿ bone screw - j7012414; 574102030 ¿ femoral stem ¿ 3037914; 00877503602 ¿ biolox ceramic head ¿ 3077143.Customer has indicated that the product will not be returned to zimmer biomet for the investigation as the product location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.It was reported a patient underwent revision approximately 1 month postop due to a deep staph aureus infection, a common skin bacteria.During the investigation process a review of the sterile certifications were reviewed and found to be conforming with no applicable deviations.Devices were verified to have gone through acceptable sterilization process following iso/aami/astm & eu published guidelines.There are multiple factors that may contribute to an infection that are outside the control of zimmer biomet, such as external factors, i.E.Hospital/surgical environment, provider related risk factors, and/or patient comorbidities/risk factors.As there are no indications of a product or process issues identified affecting implant safety or effectiveness, therefore implanted products are not identified as the source or contributing to the reported infection.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|