BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER; INTRAVASCULAR CATHETER
|
Back to Search Results |
|
Catalog Number 38833314 |
Device Problems
Break (1069); Material Separation (1562)
|
Patient Problems
Bruise/Contusion (1754); Rupture (2208)
|
Event Date 02/21/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the bd insyte¿ autoguard¿ shielded iv catheter broke.The following information was provided by the initial reporter, translated from (b)(6) to english: when introducing the catheter and pulling the mandrel, it tore the catheter, breaking the veins, leaving catheter on one side and the mandrel on the other.When a catheter was taken out of the package and turned, the mandrel broke the catheter in my hand.
|
|
Manufacturer Narrative
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Manufacturer Narrative
|
H6: investigation summary: there was no sample or photo available to bd for evaluation.Therefore, bd was unable to perform a thorough investigation to verify the reported issue.Since, an investigation could not be performed bd was unable to determine a possible root cause.A review of the device history record was performed and no quality issues were found during production.
|
|
Event Description
|
It was reported that the bd insyte¿ autoguard¿ shielded iv catheter broke.The following information was provided by the initial reporter, translated from portuguese to english: when introducing the catheter and pulling the mandrel, it tore the catheter, breaking the veins, leaving catheter on one side and the mandrel on the other.When a catheter was taken out of the package and turned, the mandrel broke the catheter in my hand.
|
|