| Model Number |
DBMNI1601 |
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Sample in unavailable for evaluation.If additional information is provided in the future, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
Bone marrow needle cant handle pressure necessary to penetrate bone, kinks and bends slips away from penetration site does not cut sample property and shatters the bone same as (b)(4).
|
| |
|
Event or Problem Description
|
|
Bone marrow needle cant handle pressure necessary to penetrate bone, kinks and bends slips away from penetration site does not cut sample property and shatters the bone.Same as (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the batch records confirmed that the product was manufactured according to specification and no deviations or anomalies were found.According to the per, no device was available to be returned for evaluation.However, an image was provided that showed that a used device was bent and the complaint was confirmed.Without the actual device to evaluate, a definitive root cause evaluation could not be accomplished.Since an evaluation of the device was not performed, no corrective action was taken.
|
| |
|