• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FLEXIVA PULSE; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FLEXIVA PULSE; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number M006L8405911
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
Note: this report pertains to one of three devices that were used in the same procedure.Refer to manufacturer report number this mfr.Report for the first flexiva pulse laser fiber, mfr report # 3005099803-2022-02946 for the second flexiva pulse laser fiber & mfr report # 3005099803-2022-02947 for the third flexiva pulse laser fiber.It was reported to boston scientific corporation on (b)(6) 2022 that three flexiva pulse 242 laser fibers were used in a percutaneous nephrolithotomy (pcnl) procedure for a stone treatment in the kidney performed on (b)(6) 2022.The laser unit used was a lumenis p120 laser console.During the procedure, when the physician plugged in each of the laser fibers and started firing, the device started smoking.The fiber connector was noted to be hot to the touch.There was no reported burn injury to the user or to the patient.The procedure was completed with a flexiva pulse 365 laser fiber.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIVA PULSE
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14549836
Report Number3005099803-2022-02945
Device Sequence Number17382637
Product Code GEX
UDI-Device Identifier08714729978626
UDI-Public08714729978626
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K210925
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/06/2025
Device Model NumberM006L8405911
Device Catalogue Number2130-01
Device Lot Number0028640672
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/06/2022
Initial Report FDA Received Date05/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-