BOSTON SCIENTIFIC CORPORATION EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7132
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that device detachment occurred.During unpacking of a 2.50mm x 12mm emerge balloon catheter, the balloon tip was bent and the device was noted to be fractured.
 
Additional Manufacturer Narrative
Date of event: used the first date of the month of the aware date as no event date was provided.
 
Event or Problem Description
It was reported that device detachment occurred.During unpacking of a 2.50mm x 12mm emerge balloon catheter, the balloon tip was bent and the device was noted to be fractured.
 
Additional Manufacturer Narrative
B3 date of event: used the first date of the month of the aware date as no event date was provided.The returned product consisted of an emerge mr balloon catheter from boston scientific batch 28840085.The device was microscopically and visually examined.There was contrast in the inflation lumen and the balloon was tightly folded.The tip of the device was damaged and prolapsed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14550617
Report Number2134265-2022-06067
Device Sequence Number6774231
Product Code LOX
UDI-Device Identifier08714729806103
UDI-Public08714729806103
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K113220
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/06/2024
Device Model Number7132
Device Catalogue Number7132
Device Lot Number0028840085
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/11/2022
Supplement Date Received by Manufacturer06/16/2022
Initial Report FDA Received Date05/31/2022
Supplement Report FDA Received Date06/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-