MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION GREENLIGHT XPS LASER SYSTEM; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION GREENLIGHT XPS LASER SYSTEM; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 0010-0210R
Medical Device Problem Code Device Displays Incorrect Message (2591)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when starting up the console, the screen displayed a goodbye message.Troubleshooting was attempted by rebooting the console; however, the message persisted.It was indicated that there was no patient involvement at the time of the event; however, the procedure was not completed due to this event.There were no clinical consequences to the patient.This event is being reported for aborted/cancelled procedure with a patient whose sedation status was unknown.
 
Event or Problem Description
It was reported that when starting up the console, the screen displayed a goodbye message.Troubleshooting was attempted by rebooting the console; however, the message persisted.It was indicated that there was no patient involvement at the time of the event; however, the procedure was not completed due to this event.There were no clinical consequences to the patient.This event is being reported for aborted/cancelled procedure with a patient whose sedation status was unknown.
 
Additional Manufacturer Narrative
The console was not returned for analysis; however, this console was serviced at the customer's facility by a field service engineer (fse).The fse confirmed that after powering up the console the "goodbye" message appeared and then the console powered off.The fse troubleshooted the system and noticed that all the fuses in the laser power supply and power distribution board were intact.After the resonator was checked, it was seen that the mode box was stuck in boot state.The fse updated the software to the resonator and that resolved the message.The fse then preemptively cleaned the fiber port and calibrated the console.The console passed all functional and electrical safety testing performed.The review completed on the dhr for this console confirmed that it met specification prior to release, and the most recent console inspection found it to be fully functional with no issues.Based on the information available, the cause that contributed to the reported courtesy error message event was most likely due to a software issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GREENLIGHT XPS LASER SYSTEM
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
alyson harris
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key14550766
Report Number2124215-2022-18951
Device Sequence Number5158165
Product Code GEX
UDI-Device Identifier00878953001906
UDI-Public00878953001906
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K092735
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0010-0210R
Device Catalogue Number0010-0210R
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/13/2022
Supplement Date Received by Manufacturer06/16/2022
Initial Report FDA Received Date05/31/2022
Supplement Report FDA Received Date07/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/13/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-