| Model Number |
404629 |
| Medical Device Problem Code |
Premature Elective Replacement Indicator (1483)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Device explanted due to eri.No adverse patient events were reported.Should additional information be received, this file will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned and analyzed.The device memory demonstrated a normal device function while the device was implanted and in service.The inspection of the device memory revealed that an elevated current was used for anti-bradycardia pacing.This indicates that high output was programmed that resulted in a faster discharge of the battery.The bench test of the icd revealed that all therapy functions were available, especially the current used for anti-bradycardia pacing was normal.In conclusion, the device was fully functional.There was no indication of a device malfunction.Analysis of the device revealed a normal battery depletion.
|
| |
|
Search Alerts/Recalls
|