It was reported to richard wolf, medwatch report (b)(4) that the "surgical team inside cvor room 3.Patient on cvor table, time out performed.Equipment passed to circulating nurse to be connected to equipment tower, surgical pencil cord, camera cord, light source cord.Surgical procedure started at 11:52 am.Approximately 3 min after procedure started at 11:55, surgeon attempted to use surgical pencil pedal.The surgical pencil pedal cord sparked twice and a flame was noticed by surgical team on surgical pencil cord.Surgeon stopped the use of the device immediately and a full assessment of the patient and room was performed.There was no harm to patient or surgical team.Surgical pencil machine, surgical pencil pedal and surgical pencil cord were exchanged.Surgical procedure continued without any other issues.".
|
|
(b)(4).This mdr is being submitted as part of an ongoing retrospective review (remediation) of prior files by (b)(4).Because the device was not returned, we are unable to investigate the device.The user is informed in the related ifu, ga-s009, about the visual and functional checks of the device.3 contraindications: warning! danger of electric shock! do not connect bipolar cables to the monopolar socket as this will cause unexpected output power.Bipolar cables are only designed for connection to the bipolar socket! contraindications directly related to the product are presently unknown.For further information refer to the latest specialized medical literature.4 combinations: caution! if the various products are combined it is absolutely necessary that the intended use and relevant technical data (working length, diameter, peak voltage etc.) correspond.Follow the instruction manuals of the products used in combination with this product.For hf applications please also follow ga-s 002 "instructions on hf application".6 checks: check the hf cable for damage before each use.Check for brittleness or cracks in the cable, worn plug, or separation of the plug from the cable.If there are any signs of damage, replace with a new cable.Check the function of the hf cable in conjunction with the hf instruments (i.E.Activate the hf instruments as described in the relevant manual).7 application: warning! hf cables with defective, brittle or cracked insulation can cause burns to the user or patient.If the electrical line breaks, an arc can occur causing burns to the user or patient or start a fire, regardless of whether the insulation is also damaged or not.Never use or repair defective hf cables! always replace them.Do not modify hf cables! the monopolar cables may be operated at a maximum recurrent peak voltage of 4000 vp and the bipolar cables at max.1000 vp.Important! when plugging and unplugging the hf cable, always hold the plug, never pull the hf cable.Do not place hf cables in parallel with other signal lines (e.G.Camera cable) to minimize the danger of interference from hf radiation.(b)(4) considers this mdr closed.Should rw receive new information, a follow-up report will be submitted.
|