Model Number BEA22-70/I20-30
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/09/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with an afx2 bifurcated stent graft and afx vela suprarenal proximal extension.Approximately one (1) year post initial procedure, the patient presented with a type 1b endoleak and a type 2 endoleak (non-device-related).The patient is currently being monitored.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with an afx2 bifurcated stent graft and afx vela suprarenal proximal extension.Approximately one (1) year post initial procedure, the patient presented with a type 1b endoleak and a type 2 endoleak (non-device-related).The patient is currently being monitored.Additional information: a re-intervention was completed.The physician elected to coil the right internal iliac artery, and implant a gore excluder (non-endologix) device in the right femoral artery.The re-intervention was successful as the endoleaks were reported to be resolved.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident-related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the type 2 endoleak (non-device related failure) and type 1b endoleak from the right limb were confirmed.This is consistent with the reported adverse event/incident.Procedure-related harms, device, user, procedure, or anatomy-relatedness of this complaint could not be determined with the medical records available for review.The final patient status remains unknown.The final patient status was not made available to endologix.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.Corrections h6: health effect, impact code- remove code 4614 h6: investigation finding codes - remove code 3233 h6: investigation conclusion codes - remove code 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key14554449
Report Number2031527-2022-00115
Device Sequence Number14723164
Product Code MIH
UDI-Device Identifier00818009015105
UDI-Public(01)00818009015105(17)230508
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 05/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/08/2023
Device Model NumberBEA22-70/I20-30
Device Lot Number2352646-005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/11/2022
Supplement Date Received by Manufacturer05/17/2022
Initial Report FDA Received Date05/31/2022
Supplement Report FDA Received Date07/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX VELA SUPRARENAL 2341320-001.
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age81 YR
Patient SexMale
-
-