| Model Number |
BEA22-70/I20-30 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.
|
| |
|
Event or Problem Description
|
|
The patient was initially treated for an abdominal aortic aneurysm (aaa) with an afx2 bifurcated stent graft and afx vela suprarenal proximal extension.Approximately one (1) year post initial procedure, the patient presented with a type 1b endoleak and a type 2 endoleak (non-device-related).The patient is currently being monitored.
|
| |
|
Event or Problem Description
|
|
The patient was initially treated for an abdominal aortic aneurysm (aaa) with an afx2 bifurcated stent graft and afx vela suprarenal proximal extension.Approximately one (1) year post initial procedure, the patient presented with a type 1b endoleak and a type 2 endoleak (non-device-related).The patient is currently being monitored.Additional information: a re-intervention was completed.The physician elected to coil the right internal iliac artery, and implant a gore excluder (non-endologix) device in the right femoral artery.The re-intervention was successful as the endoleaks were reported to be resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident-related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the type 2 endoleak (non-device related failure) and type 1b endoleak from the right limb were confirmed.This is consistent with the reported adverse event/incident.Procedure-related harms, device, user, procedure, or anatomy-relatedness of this complaint could not be determined with the medical records available for review.The final patient status remains unknown.The final patient status was not made available to endologix.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.Corrections h6: health effect, impact code- remove code 4614 h6: investigation finding codes - remove code 3233 h6: investigation conclusion codes - remove code 11.
|
| |
|