Model Number RT125
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
A distributor reported on behalf of a healthcare facility in japan that the y-piece of a rt125 infant bias flow breathing circuit was found cracked before use.There was no patient involvement.
 
Additional Manufacturer Narrative
(b)(4).The complaint rt125 infant bias flow breathing circuit is currently en route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow up report upon completion of investigation.
 
Additional Manufacturer Narrative
(b)(4).Method: the swivel of complaint rt125 infant bias flow breathing circuit was returned to fisher & paykel healthcare (f&p) in new zealand for evaluation where it was visually inspected.Results: visual inspection of device revealed a crack along one of the swivel wye ports.Conclusion: the cracking is most likely caused due to improperly mixed material.We have since contacted the supplier and arranged for them to supply the molding material with the two parts already mixed.The new pre-mixed material has now been implemented on the production line.All rt125 infant bias flow breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.The user instructions that accompany the rt125 infant bias flow breathing circuit also state the following: "check all connections are tight before use." "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient.".
 
Event or Problem Description
A distributor reported on behalf of a healthcare facility in japan that the y-piece of a rt125 infant bias flow breathing circuit was found cracked before use.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFANT BIAS FLOW BREATHING CIRCUIT
Common Device Name
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite 300
irvine, CA 92618
9494534000
MDR Report Key14554516
Report Number9611451-2022-00505
Device Sequence Number5195843
Product Code BTT
UDI-Device Identifier09420012431141
UDI-Public(01)09420012431141(10)2100944607(11)191121
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K020332
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRT125
Device Catalogue NumberRT125
Device Lot Number2100944607
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2022
Supplement Date Received by Manufacturer07/12/2022
Initial Report FDA Received Date05/31/2022
Supplement Report FDA Received Date07/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/21/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-