Model Number DFW300
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Impairment (2138); Visual Disturbances (2140); Halo (2227)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that patient could not adjust with an intraocular lens (iol).The iol was explanted from the patient's right eye due to a refractive surprise.No allegation towards lens; issue is patient not adapting with the lens.There was no patient injury and no medical or surgical intervention required.The customer did not know how the patient was doing post-operatively (op), but thinks outcome was great.Upon follow-up, it was learnt that the patient had uncorrected visual acuity (va) fluctuated from 20/40 to 20/50 with hyperopic outcome of +0.50+0.25x135.With correction, the patient could not get better than 20/25 and was very dissatisfied with glare/halos day and night in all situations.The patient could not work at the computer due to discomfort.Gave 6 months to adapt prior to proceeding with exchange.Another johnson & johnson dcb00 lens (different model and 22.5 diopter) was implanted as a replacement.Refractive information was not available as it was only one week post-op.No further information was provided.
 
Additional Manufacturer Narrative
Date of event: exact date unknown/not provided.Best estimate date is (b)(6) 2022.Device evaluated by mfr: the device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Correction: section b5: in the initial mdr, the replacement lens was indicated to be of dcb00 lens (different model and 22.5 diopter).However, the correct replacement lens was another johnson & johnson lens, model diu225 19.0 diopter.This supplemental filing is to correct the replacement lens information along with the product evaluation results.Additional info: section d9: device available for evaluation: yes.Section d9: returned to manufacturer on: jun 22, 2022.Section h3: device evaluated by manufacturer: yes.Device evaluation: the complaint lens was received loose inside of a bag.Visual inspection under magnification revealed that the lens was received cut in half.The lens was cleaned and no issues that could contribute to visual disturbances could be identified.Based on the return condition of the lens, no further product evaluation could be performed.Conclusion: the reported complaint issue was not observed.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14554978
Report Number3012236936-2022-01404
Device Sequence Number14726500
Product Code MFK
UDI-Device Identifier05050474752375
UDI-Public(01)05050474752375(17)240301
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/01/2024
Device Model NumberDFW300
Device Catalogue NumberDFW300U165
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/18/2022
Supplement Date Received by Manufacturer08/18/2022
Initial Report FDA Received Date05/31/2022
Supplement Report FDA Received Date08/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient Weight79 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-