| Model Number |
142122-14 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pain (1994); Restenosis (4576)
|
| Date of Event |
07/20/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient was treated as part of the (b)(6) study on (b)(6) 2021 to treat a restenotic lesion in the sfa proximal third of the right leg.An event of restenosis of the treated segment (target lesion) was identified on (b)(6) 2021.It was reported as not related to the device or procedure but due to a worsening of the patient condition.Doppler ultrasound on (b)(6) 2021 showed severe stenosis of right sfa stents.A target lesion revascularisation (tlr) was conducted on the (b)(6) 2021 and involved bare metal stent, drug coated balloon/drug eluting balloon (dcb/deb) and laser atherectomy.The stent was placed in mid sfa (outside of target lesion).The outcome of the event is reported as resolved/recovered.The device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
There was no reported device malfunction and the device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.The reported patient effects of restenosis is listed in the biomimics 3d instructions for use and are known patient effects of peripheral stenting procedures.There is no indication of a product quality issue with respect to the design, manufacture or labelling of the device.If further information regarding this event becomes available, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
There was no reported device malfunction and the device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.The reported patient effects of restenosis is listed in the biomimics 3d instructions for use and are known patient effects of peripheral stenting procedures.Section b.5.Has been updated with additional information received.There is no indication of a product quality issue with respect to the design, manufacture or labelling of the device.If further information regarding this event becomes available, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The patient was treated as part of the mimics 3d usa post-market observational study on (b)(6) 2021 to treat a restenotic lesion in the sfa proximal third of the right leg.A retrograde approach was used and the lesion was prepared using standard balloon angioplasty and atherectomy.The lesion was post dilated with a drug coated balloon.An event of restenosis of the treated segment (target lesion) was identified on (b)(6) 2021.It was reported as "not related" to the device or procedure but due to a worsening of the patient condition.The device relationship was later updated to "possibly related" to the device on (b)(6) 2022.Doppler ultrasound on (b)(6) 2121 showed severe stenosis of right sfa stents.A target lesion revascularisation (tlr) was conducted on (b)(6) 2021 and involved bare metal stent, drug coated balloon/drug eluting balloon (dcb/deb) and laser atherectomy.The stent was placed in mid sfa (outside of target lesion).The outcome of the event is reported as resolved/recovered.The device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
There was no reported device malfunction and the device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.The reported patient effects of restenosis leading to intervention and claudication/leg pain are listed in the biomimics 3d instructions for use and are known patient effects of peripheral stenting procedures.There is no indication of a product quality issue with respect to the design, manufacture or labelling of the device.If further information regarding this event becomes available, a follow-up report will be submitted.Sections b.3., b.5, b.7., d.3., d.4., d.10, e.1., g.1., g.6., h.6.And h.10.Have been updated.
|
| |
|
Event or Problem Description
|
|
The patient was treated as part of the mimics 3d usa post-market observational study on (b)(6) 2021.A 6.0 x 125 mm biomimics 3d stent (bm3d) was used to treat a restenotic lesion in the superficial femoral artery (sfa) proximal third of the right leg.A retrograde approach was used and the lesion was prepared using percutaneous transluminal angioplasty (pta) balloon and atherectomy.The lesion was post-dilated with a drug coated/eluting balloon (dcb/deb).An event of restenosis of the treated segment (target lesion) was identified on (b)(6) 2021.It was reported as "possibly related" to the device and not related to the procedure.It was target lesion related duplex ultrasound (dus) on (b)(6) 2021 showed severe stenosis in the right proximal sfa.The patient was experiencing rutherford class (rc) 3 symptoms.Angiographic findings included severe stenosis in the right proximal sfa stent, and mild restenosis of the mid/distal sfa stent.Between the two stents was severe diffuse stenosis.The popliteal artery had moderate stenosis.An intervention was performed on (b)(6) 2021.Laser atherectomy was performed throughout the sfa and popliteal artery.Dcb/deb of the sfa performed, followed by stenting of the mid sfa with 6.0 x 80 mm bare metal non bm3d stent.The stent was post-dilated with plain old balloon catheter (pobc)/pta/standard balloon angioplasty.The intervention was reported as a target lesion/vessel revascularisation (tlr/tvr).The outcome was reported as resolved/recovered.The device remains implanted.
|
| |
|