MEDTRONIC MEXICO SPRINTER LEGEND RX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, Back to Search Results
Model Number SPL25012X
Medical Device Problem Codes Deflation Problem (1149); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
One sprinter legend balloon catheter was attempted to be used to treat a lesion with 90% stenosis in the distal right coronary artery (rca).The device was inspected prior to use with no issues noted.The lesion was not pre-dilated.The device did not pass through a previously deployed stent.The balloon was inflated to 12 atm for 22 seconds and then for 21 seconds.  it was reported that balloon deflation difficulties occurred and the balloon would not deflate at the lesion site.The balloon was inflated to 12 atm, an attempt was made to come down and go back up to 12 atm and then attempted to come down again.The inflation device was changed but that did not work so the balloon was then pulled back into the guide and removed from the patient. a resolute onyx stent was deployed in the distal rca during the procedure. no patient injury reported.
 
Additional Manufacturer Narrative
Product analysis #(b)(4):image confirms the presence of what appears to be the sprinter legend balloon in the distal rca ¿ identified target lesion.There appears to be air in the distal balloon cone.When comparing with the procedural notes it appears that the sprinter legend balloon was inflated to 12 atms for two inflations 3 minutes apart.This suggests that there may have been no issue with the initial deflation.But this cannot be confirmed from the images.The fact that there is air in the distal portion of the balloon in the image could suggest that the balloon may have been airlocked resulting in the deflation difficulties, but this cannot be confirmed.Additional information: no difficulties were noted to remove the protective sheath.No difficulties were noted to remove the packaging stylette.Resistance was not noted while advancing the device to the lesion.A concentration of contrast / saline of 60% contrast and 40%saline was used and then 50/50 was used.Difficulties were not noted during inflations of the device.Deflation difficulties were noted after the first inflation.The device was moved or repositioned while inflated as the balloon failed to deflate so it had to be removed partially inflated.The balloon was able to be removed when it had deflated a little.  medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Device evaluation summary: the device returned for evaluation.The balloon was deflated on device return.Contrast was not visible in the balloon.The balloon material had partially pulled over the tip and was folded in on itself.The proximal balloon bond was necked.The balloon passed negative prep.It was not possible to preform inflation or deflation testing due to the condition of the proximal bond.The transition shaft material was stretched from the hypo bond to the exchange joint.A tear was evident on the exchange joints guidewire entry port.The transition shaft material was also kinked.Kinks were evident on the distal shaft.No other deformation evident to the remainder of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINTER LEGEND RX
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY,
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key14562535
Report Number9612164-2022-02085
Device Sequence Number14985320
Product Code LOX
UDI-Device Identifier00613994902832
UDI-Public00613994902832
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P790017
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/16/2024
Device Model NumberSPL25012X
Device Catalogue NumberSPL25012X
Device Lot Number223511379
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2022
Supplement Date Received by Manufacturer06/21/2022
07/26/2022
Initial Report FDA Received Date06/01/2022
Supplement Report FDA Received Date07/15/2022
08/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-