• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPORT CLEARVUE SLIM IMPLANTABLE PORT, 6CF; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPORT CLEARVUE SLIM IMPLANTABLE PORT, 6CF; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number 5676301
Medical Device Problem Code Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the device was not returned for evaluation.One case report form was provided for review.The investigation is confirmed for the reported needle bent issue.According to the case report form, approximately three months post port deployment, patient was complained of pain upon flushing of needle.Subsequently, the port needle found to be bent and needle was replaced.Later, no pain upon flushing.The severity of the adverse event was mild and recovered/resolved.A definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.(expiration date: 02/2022).
 
Event or Problem Description
It was reported that approximately three months eight days post port placement via internal jugular vein approach on the left chest, the patient had an adverse event of pain at the port site due to needle bent.It was further reported that the needle was replaced and the pain was recovered.The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT CLEARVUE SLIM IMPLANTABLE PORT, 6CF
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14562712
Report Number3006260740-2022-02028
Device Sequence Number20104265
Product Code LJT
UDI-Device Identifier00801741110795
UDI-Public(01)00801741110795
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K122899
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Other,Study
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5676301
Device Catalogue Number5676301
Device Lot NumberREEW1107
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/11/2022
Initial Report FDA Received Date06/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
Patient Weight74 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-