| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Bacterial Infection (1735)
|
| Date of Event |
03/25/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient underwent a right knee arthroplasty revision to address a deep, staphylococcus hominis and staphylococcus epidermidis infection seventeen (17) days post-operatively.Attempts have been made, however, no additional information is available at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant devices - persona cemented cruciate retaining narrow femoral component right size 5 catalog #: 42502005802 lot #: 63717984, persona medial congruent articular surface right 10mm catalog #: 42522100310 lot #: 64557869, persona tapered cemented stem extension 14mm diameter +30mm length catalog #: 42557000114 lot #: 65264231, persona cemented all poly patella 29mm catalog #: 42540000029 lot #: 65134509.The complainant has not yet indicated whether the product will be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this patient; please see all reports associated with this event: 3007963827-2022-00154.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.During the investigation process, a review of the sterile certifications were found to be conforming with no applicable deviations.The devices were verified to have gone through acceptable sterilization guidelines.There are multiple factors that may contribute to an infection that are outside the control of zimmer biomet, such as hospital/surgical environment, provider-related risk factors, and/or patient comorbidities/risk factors.As there were no indications of a product or process issue affecting implant safety or effectiveness, the implanted products were not identified to contribute to the reported infection.Zimmer biomet has determined that this event is no longer reportable based on the aforementioned statements.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|