• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERYAN MEDICAL LTD BIOMIMICS 3D VASCULAR STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERYAN MEDICAL LTD BIOMIMICS 3D VASCULAR STENT SYSTEM Back to Search Results
Model Number 144700-05
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Restenosis (4576)
Date of Event 12/10/2021
Type of Reportable Event Serious Injury
Event or Problem Description
The patient was treated as part of the mimics 3d usa post-market observational study on (b)(6) 2021.The subject was implanted with one biomimics-3d stent (a 5.0 x 150mm) to treat a denovo lesion of the sfa middle third to distal third of the left leg.A retrograde approach was used and pre-dilatation of the target lesion was conducted with percutaneous transluminal angioplasty.Atherectomy was also performed pre-stent placement.On (b)(6) 2021, a restenosis of treated segment was identified.It was reported as "not related" to the device or procedure but to a worsening of the pre-existing condition.The intervention required a target lesion revascularisation (tlr) on (b)(6) 2022 that involved bare metal stent placement, percutaneous transluminal angioplasty / standard balloon angioplasty, laser atherectomy.The patient outcome is reported as resolved/recovered.The device remains implanted.On review of the event by veryan, it was determined to be possibly related to the device due to the target lesion relationship.
 
Additional Manufacturer Narrative
There was no reported device malfunction and the device was not returned for analysis.A review has been performed of the manufacturing certificate of conformance associated with this lot.There were no non-conformances or deviations which could be associated with this event.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.The reported patient effect of restenosis is listed in the biomimics 3d instructions for use and are known patient effects of peripheral stenting procedures.There is no indication of a product quality issue with respect to the design, manufacture or labelling of the device.If further information regarding this event becomes available, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOMIMICS 3D VASCULAR STENT SYSTEM
Common Device Name
BIOMIMICS 3D VASCULAR STENT SYSTEM
Manufacturer (Section D)
VERYAN MEDICAL LTD
unit 11 galway technology park
parkmore road
galway city, H91 V E0H
EI  H91 VE0H
Manufacturer Contact
alan mcdonagh
unit 11 galway technology park
parkmore road
galway city, H91 V-E0H
EI   H91 VE0H
MDR Report Key14564700
Report Number3011632150-2022-00033
Device Sequence Number17663523
Product Code NIP
UDI-Device Identifier05391526850503
UDI-Public(01)05391526850503(17)220617(11)201116(10)0000058640
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P180003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/17/2022
Device Model Number144700-05
Device Catalogue Number144700-05
Device Lot Number0000058640
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2022
Initial Report FDA Received Date06/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/16/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ASPIRIN; CLOPIDOGREL (PLAVIX); HEPARIN IV; HEPARIN IV; HEPARIN IV; HEPARIN IV
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexFemale
Patient Weight57 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-