• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ HEPARIN LOCK FLUSH SYRINGE; HEPARIN, VASCULAR ACCESS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ HEPARIN LOCK FLUSH SYRINGE; HEPARIN, VASCULAR ACCESS FLUSH Back to Search Results
Model Number 306424
Medical Device Problem Code Missing Information (4053)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported while using bd posiflush¿ heparin lock flush syringe 2 different syringes did not have labels.There was no report of patient impact.The following information was provided by the initial reporter: it was reported that the customer found 2 syringes without labels.A nurse in the patient's home reported 2 unlabeled syringes.
 
Additional Manufacturer Narrative
It was reported the customer found two syringes without labels.To aid in the investigation, one photo was provided for evaluation by our quality team.The photo shows two syringes with no barrel label.No other defects or imperfections were observed.This defect could occur if there was a jam during the syringe barrel labeling process inducing the symptom reported.A device history record review was completed for provided material number 306424, lot number 131596n.The review did not reveal any detected quality issues during the production of this lot that could have contributed to the reported defect.Verification of the syringe barrel labeling process was performed.The settings and alignment were correct, and the flow of products were good.To date, there have been no other similar events reported for this lot.Based on the investigation and with the photo sample analysis the symptom reported by the customer is confirmed.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ HEPARIN LOCK FLUSH SYRINGE
Common Device Name
HEPARIN, VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14564959
Report Number1911916-2022-00276
Device Sequence Number20069312
Product Code NZW
UDI-Device Identifier30382903064244
UDI-Public30382903064244
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K163591
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date11/30/2023
Device Model Number306424
Device Catalogue Number306424
Device Lot Number131596N
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/04/2022
Initial Report FDA Received Date06/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-