Model Number 0168L12
Medical Device Problem Code Decrease in Pressure (1490)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that foley catheter came off about 10 minutes after inserting into the patient.After the catheter removal, the user put water again into the balloon and it deflated when the water was left for about 5 to 10 minutes.There was no balloon rupture or tear and the patient did not have any urine stone.It was stated that hand glove would have possibly contacted to the balloon.As per the additional information via email from ibc on 18may2022, no visible damage was observed to the returned sample.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that foley catheter came off about 10 minutes after inserting into the patient.After the catheter removal, the user put water again into the balloon and it deflated when the water was left for about 5 to 10 minutes.There was no balloon rupture or tear and the patient did not have any urine stone.It was stated that hand glove would have possibly contacted to the balloon.As per the additional information via email from ibc on 18may2022, no visible damage was observed to the returned sample.
 
Additional Manufacturer Narrative
The reported event is confirmed, manufacturing related.Photo samples were submitted, however, no defects were noted based on the photos.The lubricath coude foley catheter was returned without the original packaging.No visible defects were noted on the catheter.The coude tip aligned with the coude indicator.An attempt was made to inflate the catheter balloon with 10cc di water using a luer lock syringe, however, as the balloon was being inflated the water leaked from the drainage eye preventing proper inflation.All 10cc water leaked from the drainage eye and the balloon completely deflated on it's own; lumen to lumen leak was determined to be the cause of the deflation issues.The catheter does not meet specifications as it would not pass functional leak.A potential root cause for this event could be, "operator error".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A label/pack review is not required as a review of the label could not have prevented the reported event.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRICATH® FOLEY CATHETER COUDE TIP
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key14564962
Report Number1018233-2022-04259
Device Sequence Number20069614
Product Code EZC
UDI-Device Identifier00801741017803
UDI-Public(01)00801741017803
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K910846
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Model Number0168L12
Device Catalogue Number0168L12
Device Lot NumberNGEN0766
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/12/2022
Supplement Date Received by Manufacturer09/27/2022
Initial Report FDA Received Date06/01/2022
Supplement Report FDA Received Date09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-