| Model Number |
0168L12 |
| Medical Device Problem Code |
Decrease in Pressure (1490)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that foley catheter came off about 10 minutes after inserting into the patient.After the catheter removal, the user put water again into the balloon and it deflated when the water was left for about 5 to 10 minutes.There was no balloon rupture or tear and the patient did not have any urine stone.It was stated that hand glove would have possibly contacted to the balloon.As per the additional information via email from ibc on 18may2022, no visible damage was observed to the returned sample.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that foley catheter came off about 10 minutes after inserting into the patient.After the catheter removal, the user put water again into the balloon and it deflated when the water was left for about 5 to 10 minutes.There was no balloon rupture or tear and the patient did not have any urine stone.It was stated that hand glove would have possibly contacted to the balloon.As per the additional information via email from ibc on 18may2022, no visible damage was observed to the returned sample.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event is confirmed, manufacturing related.Photo samples were submitted, however, no defects were noted based on the photos.The lubricath coude foley catheter was returned without the original packaging.No visible defects were noted on the catheter.The coude tip aligned with the coude indicator.An attempt was made to inflate the catheter balloon with 10cc di water using a luer lock syringe, however, as the balloon was being inflated the water leaked from the drainage eye preventing proper inflation.All 10cc water leaked from the drainage eye and the balloon completely deflated on it's own; lumen to lumen leak was determined to be the cause of the deflation issues.The catheter does not meet specifications as it would not pass functional leak.A potential root cause for this event could be, "operator error".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A label/pack review is not required as a review of the label could not have prevented the reported event.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|
| |
|