Model Number D11LT
Medical Device Problem Codes Gas/Air Leak (2946); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.Additional information received: does the patient have a health condition that the surgeon evaluates as aggravating the event?: no.
 
Event or Problem Description
It was reported that during a oophorectomy, the product showed air leakage in the rubber (rubber running inside the trocar).
 
Additional Manufacturer Narrative
(b)(4), date sent: 12/8/2023.D4: batch # v9628f.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the d11lt device was returned with no damage in the external components.A leak test was performed and the device was noted to leak without the test probe inserted through the device.As each device is visually inspected and functionally tested during the manufacturing process, no conclusion could be reached as to what might have caused the reported event.No conclusion could be reached as to what might have caused the reported event.The reported complaint was confirmed.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XCEL DILATING TIP TROCAR 11MM
Common Device Name
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
MDR Report Key14565010
Report Number3005075853-2022-03516
Device Sequence Number4943054
Product Code GCJ
UDI-Device Identifier10705036001560
UDI-Public10705036001560
Combination Product (Y/N)N
PMA/510(K) Number
K032676
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberD11LT
Device Catalogue NumberD11LT
Device Lot Number544A85
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/15/2022
Supplement Date Received by Manufacturer11/08/2023
Initial Report FDA Received Date06/01/2022
Supplement Report FDA Received Date12/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age39 YR
Patient SexFemale
-
-