| Model Number |
D11LT |
| Medical Device Problem Codes |
Gas/Air Leak (2946); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch #: unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.Additional information received: does the patient have a health condition that the surgeon evaluates as aggravating the event?: no.
|
| |
|
Event or Problem Description
|
|
It was reported that during a oophorectomy, the product showed air leakage in the rubber (rubber running inside the trocar).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4), date sent: 12/8/2023.D4: batch # v9628f.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the d11lt device was returned with no damage in the external components.A leak test was performed and the device was noted to leak without the test probe inserted through the device.As each device is visually inspected and functionally tested during the manufacturing process, no conclusion could be reached as to what might have caused the reported event.No conclusion could be reached as to what might have caused the reported event.The reported complaint was confirmed.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|