| Model Number |
826854 |
| Medical Device Problem Code |
No Device Output (1435)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
A facility reported a cerelink icp ventricular catheter with no pressure value after insertion and monitoring was not possible.The catheter was replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint sample was returned for evaluation: dhr - there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis - no visible damage to the catheter, sensor, or connector.The readings were acceptable, and the unit passed all testing.The complaint could not be verified through failure analysis.The root cause is undetermined and was unable to be confirmed in the complaint evaluation.
|
| |
|