• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COMPANION MEDICAL INC INPEN MMT-105NNGYNA NOVO NORDISK GRAY; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COMPANION MEDICAL INC INPEN MMT-105NNGYNA NOVO NORDISK GRAY; SYRINGE, PISTON Back to Search Results
Model Number MMT-105NNGYNA
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Hyperglycemia (1905)
Event Date 05/24/2022
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Information received by medtronic indicated that inpen screw was not moving, when the button was pressed insulin did not dispense.The blood glucose value at the time of event was 290 mg/dl.No harm requiring medical intervention was reported.Troubleshooting was successfully performed; however, the customer will discontinue use of the device.Product will be returned for analysis.
 
Manufacturer Narrative
Per visual inspection: no physical damage to cap, cartridge holder or inpen was noted.Inpen received with leadscrew 1/4 of travel.High resistance was noted when trying to retract screw.The screw is not bent.However, the inpen screw and dose nut rotate together when dispensing and retracting due to broken bond at the injection nut.Unable to perform functional testing due to broken bond at the injection nut.Several attempts were made to pair inpen, every time app displayed dose doesn't match.The inpen does not pair with commercial mobile app due to broken bond.In conclusion: broken injection bond can affect insulin delivery.The customer complaint of leadscrew anomaly was confirmed.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INPEN MMT-105NNGYNA NOVO NORDISK GRAY
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer (Section G)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer Contact
tricha miles
1223 world trade drive, 100
san diego, CA 92128
7635140379
MDR Report Key14565620
MDR Text Key299863200
Report Number3012822846-2022-00816
Device Sequence Number1
Product Code FMF
UDI-Device Identifier000010862088000351
UDI-Public(01)000010862088000351(17)230510
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/10/2023
Device Model NumberMMT-105NNGYNA
Device Catalogue NumberMMT-105NNGYNA
Device Lot NumberB0132
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received06/10/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age58 YR
Patient Weight150 KG
-
-