Model Number MMT-105NNGYNA
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Fall (1848); Hypoglycemia (1912); Convulsion, Clonic (2222)
Event Date 05/13/2022
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
Information received from medtronic indicated that the customer experienced hypoglycemia.The blood glucose value at the time of call was 27 mg/dl.The customer stated that they over bloused to correct for a snack.The customer fell down in the morning and got hospitalized for hip fracture.The customer treated the low blood glucose at the emergency room and the treatment included an intravenous infusion drip and glucagon.The customer stated that they lost the product during medical emergency.Replacement set will be issued.No further complications were reported.Product will be returned for analysis.
 
Event Description
Customer also had a seizure.Product will not be returned for analysis.
 
Manufacturer Narrative
Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.The harm code for convulsion has been updated and provided with this report in section h6.The information related to the summary has been updated and provided with this report in section b5.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INPEN MMT-105NNGYNA NOVO NORDISK GRAY
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer (Section G)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer Contact
tricha miles
1223 world trade drive, 100
san diego, CA 92128
7635140379
MDR Report Key14565707
MDR Text Key293086680
Report Number3012822846-2022-00818
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-105NNGYNA
Device Catalogue NumberMMT-105NNGYNA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received08/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other;
Patient Age32 YR
Patient Weight120 KG
-
-